Nipper 4 Articulations Bohler

SKU
13.412.2

4-joint Bohler pliers, with serrated handles for better grip, and straight parallel jaws, suitable for bone surgery.

MATERIAL: stainless steel
MEASUREMENTS: total length of 15 cm, closed width of 7.5 cm
TIP TYPE: Straight jaws
HANDLE TYPE: Serrated handle for a secure grip

€90.46 Taxes not included

INDICATIONS FOR USE OF BOHLER PLIERS:

  • Bone surgery, osteosynthesis.
  • Bone-cutting shears.


TECHNICAL SPECIFICATIONS:

  • BRAND: Dimeda
  • MATERIAL: stainless steel
  • STERILIZABLE: sterilizable product
  • WEIGHT: 100 g
  • MEASUREMENTS: total length of 15 cm, closed width of 7.5 cm
  • JAW LENGTH: 2 cm in parallel position
  • JAW WIDTH: 2 cm
  • TIP TYPE: Straight jaw
  • SPRING TYPE: 2 leaf springs
  • HANDLE TYPE: Striated handle for secure grip

PRESENTATION:

  • A plastic bag with a straight shear with 4 joints, supplied in non-sterile conditions.


STORAGE:

  • Store in a cool, dry place.
  • Protect from direct sunlight.


WARNINGS:

  • Product supplied in non-sterile conditions; it must be cleaned, disinfected, and sterilized by the user before first use and after each use.
  • Keep out of reach of children.
  • Use exclusively for the intended purposes by properly qualified personnel.
  • It is essential to perform maintenance and care of the products. Visual and functional inspection before each use.
  • Ensure thorough drying after cleaning and disinfection, paying attention to gaps and recesses.
  • Discard broken, damaged, or rusted instruments, as they cannot be reused or reprocessed.

BOHLER PLIER CLEANING AND CARE:

Initial Recommendations:

  • Perform a visual and mechanical inspection of new surgical instruments.
  • Visible dirt should be removed as soon as possible to prevent residue from drying.
  • Avoid long waiting times for instrument cleaning and disinfection.

Preliminary Manual Cleaning:

  • Initially, it should be hand washed with neutral soap.
  • In the operating room, visible dirt should be removed as soon as possible to prevent encrustation, which can lead to corrosion.
  • Avoid long waiting times for reprocessing (<1h).
  • Dry cleaning is preferable, as medical devices may corrode if submerged in solutions for long periods.
  • If dry cleaning is insufficient, they should be washed with cold water (< 40ºC), including cavities, grooves, and hard-to-reach holes, which should be washed with pressurized water. Avoid leaving unwashed areas. A plastic brush with soft bristles can be used.
  • Thoroughly dry with an absorbent cloth that leaves no residue or with compressed air.

Cleaning and Disinfection:

  • Immerse the products in a solution of distilled water with an appropriate disinfectant in an ultrasonic bath or container.
  • Follow the disinfectant manufacturer's instructions to select the wash program, temperature, and time.
  • Clean instruments with movable joints in an open position or by moving the joints to avoid "shadow areas". Follow the manufacturer's instructions for dismantling detachable instruments.
  • Secure instruments with minimum separation in the cleaning container and avoid overlap to prevent product damage from collisions.
  • After the disinfection program, rinse the products with plenty of cold distilled water, paying particular attention to cavities, grooves, and hard-to-reach holes, if possible, with pressurized water.
  • Thoroughly dry with an absorbent cloth that leaves no residue or with compressed air.

Lubrication:

  • Use Sterilit Protector Instrument (ref. 13.859.1 and 13.859.2) to protect the instruments.
  • Attach the white plastic tube included with the spray and apply the oil spray directly to the joint area or hard-to-reach areas.
  • Open and close the instrument several times to distribute the oil evenly and remove excess oil with a lint-free cloth. Then proceed to sterilize the instrument, as the oil film is steam-permeable. It will act as a lubricant and anticorrosive without leaving marks on the instruments.

Sterilization:

  • Sterilize products considering applicable national requirements.
  • The products must be sterilized in appropriate sterilization packaging.