Removable Handpiece ES6 Krafit

SKU
17.002.18

Removable handpiece model ES6 Krafit – Removable dental handpiece for Marathon micromotors. The ES6 KRAFIT is a lightweight, durable and ergonomic handpiece made of stainless steel. It reaches speeds of up to 40,000 rpm, providing precise and stable performance.

COMPATIBILITY: Marathon micromotors:

- Ref. 17.002.17

- Ref. 17.002.19

More information
€175.00 Taxes not included

ES6 KRAFIT HANDPIECE

The ES6 KRAFIT is a lightweight, durable, and ergonomic handpiece made of stainless steel. Designed to integrate with Marathon micromotors, it can reach speeds of up to 40,000 rpm, offering precise and stable operation. It is fully autoclavable, ensuring maximum hygiene and extending its service life. Its compact and balanced design provides a comfortable user experience, even during long clinical sessions.

INDICATIONS FOR USE:

  • Recommended for dental and podiatry procedures.
  • Compatible with Marathon micromotors.
  • Designed as a professional accessory for clinical tasks requiring precision.

FEATURES:
MATERIAL: Stainless steel
MAXIMUM SPEED: 40,000 rpm
STERILIZATION: Autoclavable – 4 minutes at 132 °C + 30 minutes drying
DIMENSIONS: 92 mm (length) x 20 mm (diameter)
DESIGN: Ergonomic and balanced handle
COMPATIBILITY: Marathon micromotors:

- Ref. 17.002.19

- Ref. 17.002.19


MODEL: SDE ES6

PRESENTATION: Individual box containing one SDE ES6 handpiece.

INSTRUCTIONS FOR USE:

  • After each use, disconnect the handpiece from the micromotor.
  • Clean the exterior with a soft cloth moistened with isopropyl alcohol.
  • Lubricate the handpiece using a specific spray:

- Shake the spray can.
- Insert the nozzle into the rear port.
Spray for 2 to 3 seconds.

For autoclave sterilization:

  • Place the handpiece in a sterilization pouch.
  • Seal the pouch.
  • Sterilize in an autoclave at 132 °C for 4 minutes and dry for 30 minutes.
  • After the process, keep clean and lubricated until next use.

WARNINGS:

  • Read and follow the instructions before first use.
  • Always sterilize before using on patients.
  • Do not use if abnormal noise or vibrations are detected.
  • Ensure the handpiece is completely stopped before disconnecting it.
  • Do not exceed the maximum recommended speed to avoid premature wear.