Metzenbaum Dimeda surgical scissors, made of high-quality stainless steel. Available in straight version (for superficial tissues) or curved version (for deeper and hard-to-reach tissues). Reusable, sterilizable, and latex-free.
LENGTH: 11.5 cm or 14.5 cm, depending on reference.
The Metzenbaum ScissorsDimeda are precision surgical instruments designed to cut and dissect delicate tissues, vessels, viscera, and ducts.
Their blunt/blunt tip design prevents accidental perforations.
The straight model (Ref. 13.208.21) is mainly used for superficial tissues.
The curved model (Ref. 13.208.22 and 13.208.23) is indicated for deeper planes and areas that are difficult to reach.
Made from high-quality stainless steel, they ensure durability, safety, and reliable cutting performance in surgeries and medical procedures. Supplied non-sterile; must be reprocessed before first use.
INDICATIONS FOR USE:
Blunt dissection and cutting of delicate tissues.
General, vascular, visceral, and soft tissue surgery.
Straight → superficial tissues.
Curved → deep or hard-to-reach tissues.
FEATURES:
Ref. 13.208.21 – Metzenbaum Straight 14.5 cm
Total length: 14.5 cm
Width: 5 cm
Tip length: 5.3 ± 0.1 cm
Weight: 35 ± 2 g
Tip: straight, blunt/blunt
Material: stainless steel
Brand: Dimeda
Ref. 13.208.22 – Metzenbaum Curved 14.5 cm
Total length: 14.5 cm
Width: 5 cm
Tip length: 5.0 ± 0.1 cm
Weight: 35 ± 2 g
Tip: curved, blunt/blunt
Material: stainless steel
Brand: Dimeda
Ref. 13.208.23 – Metzenbaum Curved 11.5 cm
Total length: 11.5 cm
Width: 5 cm
Tip length: 4.3 ± 0.1 cm
Weight: 24 ± 2 g
Tip: curved, blunt/blunt
Material: stainless steel
Brand: Dimeda
PRESENTATION:
1 unit per plastic bag.
Non-sterile product → sterilize before first use.
INITIAL RECOMMENDATIONS:
Visual and functional inspection before first use.
Manual washing with water and neutral soap prior to cleaning and disinfection.
Remove organic residues immediately to avoid corrosion and buildup.
Avoid long delays before reprocessing (< 1 h).
CLEANING, DISINFECTION AND STERILIZATION:
Manual cleaning: preferably dry; otherwise, wash with cold pressurized water, including grooves and cavities.
Disinfection: immerse in suitable solution with ultrasound, rinse with distilled water and dry thoroughly.
Lubrication: apply protective oil to joints before sterilization.
Steam sterilization: 15 min at 121 °C (110 kPa) or 3 min at 134 °C (210 kPa).